The AI Act lays down harmonised rules for placing AI systems on the market, putting them into service and using them in the Union. The activities before an AI system reaches the market are, as a rule, outside its scope. Under Art. 2(8), the Act does not apply to research, testing or development of AI systems and models before they are placed on the market or put into service. That exclusion does not, however, cover testing in real-world conditions, which Art. 2(8) expressly brings back within the Act's scope. For projects that develop medical AI and test it on patients, it is therefore essential to establish which studies amount to 'testing in real-world conditions' and which obligations apply at each stage. This post sets out how we answered these questions for AIRCARE, how the AI Act interacts with the rules on clinical investigations under the Medical Device Regulation (MDR), and which practical steps other projects should take.
The AI Act is modelled on EU product safety legislation under the New Legislative Framework. It regulates a product, the AI system, and follows a risk-based approach: it prohibits a limited number of practices, imposes transparency obligations on certain systems, sets specific rules for general-purpose AI models and reserves its most stringent requirements for high-risk AI systems. Those requirements fall mainly on the provider.
The AI Act identifies high-risk systems in two ways. Under Art. 6(2) and Annex III, certain uses are high-risk in themselves, such as triaging patients in emergency care. Under Art. 6(1), an AI system is high-risk where it is a product, or a safety component of a product, covered by Union harmonisation legislation listed in Annex I, such as the MDR or IVDR, and that product must undergo third-party conformity assessment.
Software is a medical device in its own right where the manufacturer intends it for a medical purpose such as diagnosis or treatment. Software that detects and characterises lesions on endoscopic images or delineates tumour margins before surgery, as AIRCARE's systems are designed to do, meets that test and therefore the first condition in Art. 6(1)(a) AI Act. Under Rule 11 of Annex VIII MDR, software providing information used for diagnostic or therapeutic decisions is at least class IIa, and class IIb or III where those decisions may cause serious or irreversible harm. From class IIa upwards, conformity assessment requires a notified body. That is the third-party assessment required by Art. 6(1)(b), and it makes such software a high-risk AI system.
The Digital Omnibus postponed the application of Art. 6(1), and with it the high-risk requirements, for Annex I products, including medical devices, to 2 August 2028. Until then, AI-enabled medical devices remain fully subject to the MDR but not to the AI Act's high-risk requirements. Devices placed on the market or put into service before that date fall within the AI Act only if they later undergo significant changes in their design (Art. 111(2)).
Before the high-risk rules come into play, a prior question arises as to whether the AI Act applies to research at all. The answer depends partly on what the research is for. Some projects develop systems intended to be placed on the market and, eventually, used on patients; others build tools used purely for scientific research, which never leave the laboratory. To support innovation and freedom of science (recital 25), the AI Act leaves both kinds of research largely outside its framework through two separate exclusions. Activities covered by either exclusion fall outside the Act altogether, so neither the classification rules nor the obligations attached to them come into play. They are often referred to together as "the research exemption", which is a source of error: Art. 2(6) turns on the purpose of the AI system, Art. 2(8) on the stage of its life cycle.
Table 1. The two research exemptions in Art. 2 AI Act
| Development-phase exemption | Scientific-use exemption | |
| Provision | Art. 2(8) AI Act | Art. 2(6) AI Act |
| What it covers | "any research, testing or development activity regarding AI systems or AI models prior to their being placed on the market or put into service" | AI systems or AI models, "including their output, specifically developed and put into service for the sole purpose of scientific research and development" |
| Stage covered | Pre-market only: design, training, validation and lab testing, even where a commercial product is planned | Development and use: the system is put into service, but only as a research tool, and is never placed on the market |
| Where it stops | "Testing in real world conditions shall not be covered by that exclusion" | Once the system is used for anything other than scientific research |
Art. 2(6) thus exempts AI that serves research alone, for as long as it is used for that purpose. Art. 2(8) exempts any AI system, including a future commercial product, but only until it reaches the market and only until testing in real-world conditions begins. For product-oriented projects, the decisive question is whether, and when, a study crosses that line.
Art. 3(57) defines testing in real-world conditions as "the temporary testing of an AI system for its intended purpose in real-world conditions outside a laboratory or otherwise simulated environment, with a view to gathering reliable and robust data and to assessing and verifying the conformity of the AI system with the requirements of this Regulation and it does not qualify as placing the AI system on the market or putting it into service within the meaning of this Regulation, provided that all the conditions laid down in Article 57 or 60 are fulfilled." The definition has a contextual element (temporary testing, for the intended purpose, outside a laboratory or simulated environment) and a purpose element (gathering reliable data and verifying conformity with the AI Act). The contextual element draws the line between development and testing: what matters is whether the system meets the real setting in which it is meant to work, with real users, real inputs and real people affected, rather than a controlled reproduction of it. C
rossing that line does not mean the system must already comply with the AI Act as a whole. What is regulated at this stage is the testing itself, under Art. 57 (within a regulatory sandbox) or Art. 60 (outside one). Outside a sandbox, the conditions include a testing plan approved by the market surveillance authority, registration, a six-month limit (extendable once), informed consent under Art. 61, qualified oversight, outputs that can be reversed and disregarded, and incident reporting. Testing that does not meet them risks being treated as putting the system into service.
AIRCARE (AI-augmented Robotics for CAncer point of caRE) is a Horizon Europe Innovation Action running from July 2024 to June 2028, coordinated by the Istituto Italiano di Tecnologia with eleven partners. It targets cancers of the mouth, throat and voice box, which are found mainly through endoscopy. Reading those images is subjective and depends on the specialist's experience, and AIRCARE's premise is that AI can extract more objective information from them. Its studies sit at different points between the lab and the patient:
• Retrospective dataset study. Archived recordings of routine endoscopies are annotated by expert surgeons (lesion, edges, pathology result) and used to train the models. No patient is examined or treated.
• Ex vivo bioimpedance study. A pen-like sensor measures how tumour tissue already removed in routine surgery responds to a very small electrical current. The measurements are matched with pathology to build tissue-classification models. The probe never touches the patient.
• Shadow-mode clinical study. The AI runs live on the endoscopy video of real patients, in the outpatient clinic and the operating room, detecting suspicious areas, characterising lesions and outlining tumour edges. Its outputs are recorded but hidden from the physician, so care follows standard practice, and are later compared with pathology.
None of these studies can rely on Art. 2(6) exemption. The project develops AI systems with later commercialisation in mind, so they are not developed "for the sole purpose of scientific research and development", even while tested in a research setting. That leaves the Art. 2(8) exemption to consider. The first two studies clearly fall within it: one runs offline on archived data, the other tests the probe on excised tissue rather than in the intraoperative conditions it is designed for. The shadow-mode study is the harder case. Because the AI does not influence care, it is tempting to treat it as development, but measured against Art. 3(57) it is likely to meet both elements.
The study is limited in time, the system performs the function it is designed for, namely detecting, characterising and outlining lesions, and it does so on live patients in the outpatient clinic and in the operating room (the contextual element). As for the purpose element, the system is being developed with a view to CE marking, and the data the study generates will be needed to demonstrate its conformity with the AI Act once it reaches the market. Both elements are discussed in more detail below.
AIRCARE's studies therefore illustrate the three stages the AI Act distinguishes: (1) research, development and testing outside real-world conditions, which remain within Art. 2(8), as do the first two studies; (2) testing in real-world conditions, which must comply with Art. 57 or 60 or, for Annex I products such as medical devices, with the sectoral testing rules preserved by Art. 60(1), third subparagraph, as the shadow-mode study must; and (3) placing on the market or putting into service, which triggers the full high-risk requirements. Stages 2 and 3 also depend on timing, since both regimes apply only to high-risk AI systems, and AI-enabled medical devices qualify as such only from 2 August 2028.
In a shadow-mode (or silent-mode) study, the system runs in the background on real cases but its outputs are not shown to clinicians. It is often assumed that such a study carries few legal obligations. Under the AI Act, that assumption is difficult to sustain. In our view, the study constitutes testing in real-world conditions within the meaning of Art. 3(57).
The contextual element is met on all three counts. First, the testing is temporary: it runs for the period set in the protocol and ends once enough cases have been compared with pathology. Second, it concerns the system's intended purpose, which is to detect, characterise and outline lesions on endoscopic video, exactly what the system does in the study. It could be objected that the intended purpose includes "the specific context and conditions of use" (Art. 3(12)), which for a diagnostic system means supporting the clinician's decision. In our view, that reads too much into the definition. Art. 3(57) requires testing for the intended purpose, not use for that purpose. The patients, inputs, clinical setting and workflow are those of routine care, and only the final step, the clinician acting on the output, is held back. The AI Act itself suggests that step is not essential, since outputs produced during real-world testing must be capable of being "effectively reversed and disregarded" (Art. 60(4)(k)); shadow mode simply applies that safeguard from the outset. Third, the testing takes place in real-world conditions. Patients, anatomy, image quality, lighting and workflow are neither selected nor reproduced for the test. Shadow mode leaves untested how clinicians interact with the output, which limits the evidence produced but does not turn the clinic into a simulated environment.
On the purpose element, it could be argued that a study designed to produce scientific evidence, rather than to prepare a conformity assessment, falls outside the definition. That is difficult to sustain where, as in AIRCARE, the system is being developed towards CE marking. Once Art. 6(1) applies, such a device can only be placed on the market after an MDR conformity assessment that also covers the AI Act's high-risk requirements (Art. 43(3)), on the basis of a single set of technical documentation (Art. 11(2)) showing how the system was tested and how accurate and robust it is. Real-patient performance data are precisely that evidence. Purpose should be judged objectively, by what the data will be used for, not by how the protocol describes it. Withholding outputs reduces the risk to patients, but it does not change the regulatory status of the study.
A similar assumption is often made under the MDR, namely that hiding the outputs keeps the study outside the rules on clinical investigations. Art. 2(45) MDR defines a clinical investigation as "any systematic investigation involving one or more human subjects, undertaken to assess the safety or performance of a device". The notion turns on what a study assesses, not on its stage: the MDCG treats first-in-human and feasibility studies as pilot-stage clinical investigations and recommends the Art. 62 route in case of doubt. The shadow-mode study is plainly systematic, following a protocol with predefined endpoints, and undertaken to assess performance.
The only arguable element is whether it "involves" human subjects. The MDCG accepts that a retrospective analysis does not, because exposure to the device precedes the study, management is unaffected and no additional risk arises (Q18). Extending that reasoning to shadow mode is difficult. Patients are enrolled prospectively, the device runs on their examinations and surgery as they take place, and its outputs are generated for the study. The absence of influence on care is also less certain: blinding reduces, but cannot entirely exclude, effects on care, for instance through incidental findings that cannot ethically be ignored, or through separation arrangements that do not always work in practice. Shadow mode is therefore best seen as a risk-reduction measure rather than a reason to step outside the MDR, whose clinical investigation framework is designed to manage precisely this kind of residual risk. Where doubts remain, sponsors should document their assessment, follow Art. 62 for early-stage studies and, if needed, consult the competent authority. This dual qualification is consistent with the joint AIB and MDCG guidance, which states that a clinical investigation of high-risk medical device AI under the MDR constitutes real-world testing under the AI Act (Q25). Although not binding, it is the clearest indication so far of how regulators read the interface.
When an AI-enabled medical device that qualifies as high-risk under Art. 6(1) is tested in shadow mode, the testing is governed primarily by the MDR rather than by a separate AI Act procedure. Three steps lead to that conclusion.
The guidance predates the Digital Omnibus, which extended Art. 60 to high-risk AI systems covered by Section A of Annex I, including medical devices. In our view its reasoning remains valid, but the point is not settled, and national authorities or notified bodies could read Art. 60(1) differently. The Commission has acknowledged the need for guidance on the research exemptions, including for medical device development, and undertook to prioritise it. The cleaner solution would be legislative: either Art. 3(57) should expressly cover testing under the sectoral law preserved by Art. 60(1), or the clinical investigation procedure should integrate the AI Act requirements.
In practice, the question becomes pressing only on 2 August 2028. Until then, AI-enabled medical devices are not high-risk AI systems, so no AI Act high-risk or Art. 60 obligations apply to their testing. For AIRCARE, which ends in June 2028, the shadow-mode study runs under the MDR, the GDPR and national law. The issue will arise for the pivotal investigations that follow, which will likely run past that date and beyond six months. Pending clarification, the safer course is the cumulative reading: design the clinical investigation plan so that it also meets Art. 60(4) AI Act, namely the elements of a real-world testing plan, informed consent covering the information required by Art. 61, effective oversight, reversibility of outputs and registration. It is also advisable to consult the market surveillance authority early, to consider an AI regulatory sandbox for novel devices, and to follow upcoming guidance and legislative changes.
Within several months, two initiatives pulled the AI Act and MDR interface in opposite directions. In December 2025, the Commission's targeted revision of the MDR and IVDR proposed moving both regulations to Section B of Annex I, where the AI Act would apply only residually. The Digital Omnibus, in force since 27 July 2026, kept medical devices in Section A, so they will qualify as high-risk AI systems, but postponed the application of Art. 6(1), and with it the high-risk obligations, to 2 August 2028 and introduced an equivalence mechanism instead of an exemption. The MDR revision is still pending, so the Section B option is not formally off the table.
Table 2. The MDR/IVDR targeted revision and the Digital Omnibus on AI compared
| MDR/IVDR targeted revision | Digital Omnibus on AI | |
| Instrument | COM(2025) 1023, 16 December 2025 | Regulation (EU) 2026/1744, 8 July 2026 |
| Status | Pending | In force |
| Medical devices | Would move to Annex I, Section B | Stay in Annex I, Section A |
| AI Act applies | Only residually (Art. 2(2)): Art. 6(1), Art. 60a and Arts. 102 to 112 | In full, from 2 August 2028, subject to the equivalence mechanism of new Art. 2(13) |
Under new Art. 2(13), the requirements in Arts. 9 to 15 and 17 to 25 (the substantive high-risk requirements and the obligations of operators along the value chain) "may be limited, where and to the extent that" Section A legislation "lays down requirements or obligations providing an equivalent or higher level of protection of health, safety or fundamental rights", and the limitation "does not reduce the overall level of protection provided for by this Regulation". The Commission must adopt the delegated acts by 2 August 2027; none has been adopted yet. Any limitation will matter at CE marking rather than during testing, as Art. 2(13) does not extend to the real-world testing rules in Arts. 57 to 61. Two safeguards central to clinical AI are unlikely to benefit from it.
Table 3. Two AI Act safeguards unlikely to be covered by equivalence4
| AI Act safeguard | Equivalent in the MDR? | Consequence |
| Art. 14 human oversight | Largely no. MDR usability and use-error requirements (Annex I, section 5 MDR) overlap, but the MDR does not require that users can interpret, disregard or override outputs, or address automation bias (Art. 14(4) AI Act) | Design for oversight stays an AI Act requirement |
| Art. 10 data and data governance | Unlikely. The MDR addresses the software lifecycle and verification and validation (Annex I, section 17.2 MDR), not the representativeness of datasets or the examination of bias (Art. 10(2)(f) to (g) and (3) AI Act) | Training, validation and testing data must meet the AI Act standard |
Risk management (Art. 9) and quality management (Art. 17) are likelier candidates for relief, since the MDR already lays down comparable requirements in Annex I and Art. 10(9). The decision on equivalence has, however, been left to the Commission rather than taken by the co-legislators. That choice is open to question: under Art. 290 TFEU a delegated act may supplement or amend only non-essential elements of the basic act, whereas deciding what counts as equivalent effectively determines which AI Act safeguards continue to apply to medical devices.
A study that runs an AI system prospectively on patients is likely to qualify both as testing in real-world conditions under the AI Act and as a clinical investigation under the MDR. Projects developing AI-enabled medical devices should map the rules applicable over the project's lifetime, establish which exemptions cover their activities and where they end, and prepare from the outset for the obligations that apply once the system is placed on the market. Most of the steps below must be built in at protocol design.
Anushka Sachdev participates in the AIRCARE project. AIRCARE is funded by the European Union under the Horizon Europe Programme, Grant Agreement No. 101137426. Marta Wilinska participates in the AI4LUNGS project, funded by the European Union under the Horizon Europe Programme, Grant Agreement No. 101080756. The Health and Digital Executive Agency (HaDEA) is the granting authority for both projects. Views and opinions expressed are those of the authors only and do not necessarily reflect those of the European Union or the granting authority. Neither the European Union nor the granting authority can be held responsible for them.